T E C H N I C A L D A T A | ||||
Talyrond 30 series | ||||
| TALYROND 30 | TALYROND 31 | ||
Worktable & Spindle | Worktable diameter | 125mm (4.94IN) | 125mm (4.94in) | |
Range of manual centring | ±1.25mm (±0.054in) | ±1.25mm (±0.054in) | ||
Range of manual leveling | ±30 arc minutes | ±30 arc minutes | ||
Height of neutral plane | 51mm (2in) | 51mm (2in) | ||
Speed of rotation | 6rpm clockwise nominal | 6rpm clockwise nominal | ||
Spindle roundness limit of error (concentric load)* | ±0.025mm + 0.00025mm/mm | (±0.975min + 0.0099min/in) | ||
System roundness limit of error (concentric load)* | ±0.04mm + 0.00025mm/mm | (±1.57min + 0.0099min/in) | ||
System roundness limit of error (eccentric load)* | ±0.04mm + 0.0005mm/mm | (±1.57min + 0.02min/in) | ||
Axial limit of error | 0.025mm (0.975min) | 0.025mm (0.975min) | ||
Note *: Departure from the least squares circle at 6rpm with 0-50 upr filter (departure may be inward or outward). | ||||
Cylindricity Reference Unit | Straightness/parallelism error (over total column 225mm)** | -3mm(118min) | ||
Parallelism (orthogonal to measuring plane) | 30mm over 225mm (0.001in over 9in) | |||
Note *: Straightness/parallelism specification based on calculations from a series of radial traces using an LS circle. | ||||
Gauge Range & Resolution | Maximum range | ±1mm (±0.039in) | ±1mm (±0.039in) | |
Resolution at maximum range | 0.06mm (2.5min) | 0.06mm (2.5min) | ||
Minimum range | ±200mm (±0.008in) | ±200mm (±0.008in) | ||
Resolution at min range | 0.012mm (0.5min) | 0.012mm (0.5min) | ||
Measuring Capacity | Max diameter | | ||
Primary gauge | 200mm (8in) | 370mm (15in) | ||
Secondary gauge | 200mm (8in) | 200mm (8in) | ||
Max height | | |||
Primary gauge | 200mm (8in) | 225mm (9in) | ||
Secondary gauge (optional) | 200mm (8in) | 200mm (8in) | ||
Max weight (balanced load) | 12kg (25lb) | 12kg (25lb) | ||
Filter options | Phase corrected 2CR, Gaussian and bandpass filtering available by menu selection. Filtering is operator selectable from 1-15upr, 1-50upr, 1-500upr, 15-150upr, 15-500upr, and user selected. | |||
Electrical Supply | Alternating supply, single phase with earth (3 wire system) | |||
Voltage | 90-260v | 90-260v | ||
Frequency | 47-63Hz. | 47-63Hz. | ||
Consumption (total system) | 250VA max, 160w | 250VA max, 160w | ||
Environmental Conditions | Temperature: | Operating*** Storage | 10-35°C | 10-35°C |
| | -10 to + 50°C | -10 to + 50°C | |
Temp/time gradient | Less than 2°C/hour | Less than 2°C/hour | ||
Humidity | Operating Storage | 30% to 80% relative, non-condensing | ||
| 10% to 90% relative non-condensing | |||
Free air flow rate | 1.0m/sec maximum steady | | ||
*** For optimum accuracy, the instrument should be operated within ±2°C of the temperature at which it was calibrated | ||||
Air Source Requirements | Maximum source pressure | 8.1bar (120psi) | 8.1bar (120psi) | |
Minimum source pressure | 5.4bar (80psi) | 5.4bar (80psi) | ||
Air consumption | 0.037cu.m/min (1.3scfm) | 0.037cu.m/min (1.3scfm) | ||
Operating pressure | 4.1bar (60psi) | 4.1bar (60psi) | ||
Filtering | 5mm (200min) | 5mm (200min) | ||
Moisture content – dew point | 2°C | 2°C | ||
Weights and Dimensions | Main instrument: overall weight | 38kg (85lb) | 62kg (136lb) | |
Overall length | 330mm (13in) | 820mm (32in) | ||
Overall width | 280mm (11in) | 343mm (13.5in) | ||
PC: overall weight: | 30kg (66lb) | |||
Talyrond 30 and 31 conform to the European Electromagnetic Compatibility regulations EN50081-1 and EN50082-1. Note: To allow for supply fluctuations, the preferred air pressure is 80-120lbf/in2(5.4 to 8.1 bar) however where a constant supply pressure can be maintained, a minimum air pressure of 50lbf/in2 (3.4bar) is acceptable. Talyrond 30 Series instruments conform to all applicable requirements of BS EN 50081-1 (1992) and BS EN 50082-1 (1992) and comply with the requirements of the EMC Directive 89/336/EEC. Roundness, concentricity, eccentricity and run out measurements can be made with reference to Least Square Circle, Minimum Zone Circle, Maximum Inscribed Circle or Minimum Circumscribed Circle, flatness and square ness to Least Squares and Minimum Zone. | ||||